Journal of Graphics
Previous Articles Next Articles
Online:
Published:
Abstract: A method for suppressing motion artifacts due to cyclic cardiac motion existing in continuous pullback coronary artery intravascular ultrasound (IVUS) image sequence acquired in vivo is proposed. Firstly, rigid motion parameters of cross-sectional vessel wall in tomographic image sequence are quantitatively estimated based on comprehensive analysis of both mechanism and expressions of motion artifacts. Then, dynamic and geometrical components of rigid motion parameters are separated with spectrum analysis technique. Finally, geometrical components of vessel wall motion in successive tomographic frames are quantitatively compensated to implement suppression of motion artifacts. Neither special ECG-gating acquisition device nor ECG signals are required. It is not required to discard useful frames to guarantee the integrity of the image data set. The validity has been demonstrated with computer-simulated and clinically acquired image data.
Key words: coronary artery, intravascular ultrasound (IVUS), motion artifact, cardiac motion, suppression
Sun Zheng. A method for suppressing motion artifacts in IVUS image sequence[J]. Journal of Graphics.